Experiences with interlaboratory GPC experiments

Rudolf Bruessau

Macromolecular Symposia · 1996 · 40 citations · 0 references

DOIFull text

Open access

Concepts

Abstract

Abstract Based on some round robin experiments organized in the last years by a DIN working group and by the IUPAC working party IV.2.2, the paper discusses measures for getting a better interlaboratory reproducibility of GPC results. The most important points are: ‐ selecting columns according to their separation selectivity and range ‐ avoiding artifacts in the distribution curves, caused by gaps in the pore volume distribution ‐ avoiding adsorption and memory effects ‐ calibration: using calibration standards of high quality selecting suitable calibration points checking the influence of the injected concentration ‐ selecting a common detector ‐ data treatment, especially finding the correct baseline ‐ preparing a detailed data report according to a uniform schedule.