Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · 2005 · 11 citations · 0 references
EngineeringBioengineeringBiomedical EngineeringBiomedical ImagingBiomedical OpticsBiophotonicsInterferometryMedical ImagingOptical CoherenceOptical Coherence TomographyOptical ImagingOptical TomographyMach-zehnder InterferometersOptical Coherence Tomography InstrumentationPolarization In Tissue OpticsPolarization-sensitive DetectorsPolarization-sensitive ImagingPolarization-sensitive Optical Coherence TomographyPolarization-sensitive TomographySuperluminescent Diodes
We present, to the best of our knowledge, the first ultrahigh resolution, polarization sensitive OCT images of human tissue in vivo. The system is based on a Mach Zehnder interferometer and a transversal scanning of the sample. A broadband superluminescent diode with a center wavelength of 897nm and a bandwidth (FWHM) of 155nm was used. The depth resolution of the system was measured with ~ 4 micrometer in air. The actual scanning speed of 1000 transversal lines per second enables the acquisition of an image (B-scan) consisting of 1600(x) x 1000(z) pixels in 1 second. The whole signal is recorded by a polarization sensitive detection unit at the interferometer exit which enables a phase resolved measurement. From the recorded data we were able to obtain backscattered intensity, retardation and cumulative fast axis orientation of the sample. Images of these parameters obtained from a technical sample and from a human cornea in vivo are presented.