Journal of Speech Language and Hearing Research · 2003 · 122 citations · 10 references
Natural SciencesSpeech-language PathologyAcousticsAssistive Listening DevicesInstrumentationSpeech AcousticsVocal AssessmentVocal HealthVoice ControlAccelerometer MonitoringAccelerometer SensorsRecordingsSound Pressure LevelSound Pressure Level AnalysisSound Pressure Level MeasurementVocal Function AssessmentVoice Recording
This research note describes the design and testing of a device for unobtrusive, long-term ambulatory monitoring of voice use, named the Portable Vocal Accumulator (PVA). The PVA contains a digital signal processor for analyzing input from a neck-placed miniature accelerometer. During its development, accelerometer recordings were obtained from 99 participants with normal or dysphonic voices. The recordings were used to (a) test the specifications and capabilities of the PVA for monitoring normal and dysphonic voices and (b) explore potentially useful displays for the large quantity of data generated by long-term monitoring. The current prototype PVA is pocket-sized (12 x 8.5 x 2 cm), lightweight (200 g), and capable of sampling 11 hr of voice-use data, including estimates of fundamental frequency, sound pressure level, and phonation duration.
10
Voice and Articulation Drillbook
C. E. Seashore · Music Educators Journal · 1940 · 897 citations
Vocal Load as Measured by the Voice Accumulator
Romain Buekers, E. Bierens, Herman Kingma et al. · Folia Phoniatrica et Logopaedica · 2009 · 95 citations
Medicine, Rehabilitation Medicine, Speech-language Pathology +17