Concepedia

Publication | Closed Access

Two-color pump-probe laser spectroscopy instrument with picosecond time-resolved electronic delay and extended scan range

52

Citations

48

References

2005

Year

Abstract

An electronically delayed two-color pump-probe instrument was developed using two synchronized laser systems. The instrument has picosecond time resolution and can perform scans over hundreds of nanoseconds without the beam divergence and walk-off effects that occur using standard spatial delay systems. A unique picosecond Ti:sapphire regenerative amplifier was also constructed without the need for pulse stretching and compressing optics. The picosecond regenerative amplifier has a broad wavelength tuning range, which suggests that it will make a significant contribution to two-color pump-probe experiments. To test this instrument we studied the rotational correlation relaxation of myoglobin (τr=8.2±0.5ns) in water as well as the geminate rebinding kinetics of oxygen to myoglobin (kg1=1.7×1011s−1, kg2=3.4×107s−1). The results are consistent with, and improve upon, previous studies.

References

YearCitations

Page 1