Optics Express · 2014 · 120 citations · 15 references
EngineeringElectrical EngineeringInstrumentationMetrologyOpticsAbsorption SpectroscopyDistance MeasurementInterferometryMeasurement UncertaintyOptical InstrumentationOptical MetrologyGas AbsorptionMeasurement System DesignMeasurement System ImplementationOptical Distance MeasurementOptical InterferometryOptical Measurement SystemsOptical Path Difference MeasurementOptical Path Length
The gas absorption cell serves as the length reference, traceable to international standards via its absorption feature frequencies. We present an implementation of an absolute distance measurement system based on frequency scanning interferometry (FSI). Dynamic FSI employs two frequency‑scanning lasers, a gas absorption cell, and a reference interferometer to determine the optical path length difference of one or more measurement interferometers, with variations sampled at 2.77 MHz. The system measures distances from 0.2 m to 20 m with a combined relative uncertainty of 0.41 × 10⁻⁶ (k = 2) and can resolve OPD changes to 40 nm within a scan.
We present an implementation of an absolute distance measurement system which uses frequency scanning interferometry (FSI). The technique, referred to as dynamic FSI, uses two frequency scanning lasers, a gas absorption cell and a reference interferometer to determine the unknown optical path length difference (OPD) of one or many measurement interferometers. The gas absorption cell is the length reference for the measurement system and is traceable to international standards through knowledge of the frequencies of its absorption features. The OPD of the measurement interferometers can vary during the measurement and the variation is measured at the sampling rate of the system (2.77 MHz in the system described here). The system is shown to measure distances from 0.2 m to 20 m with a combined relative uncertainty of 0.41 × 10⁻⁶ at the two sigma level (k = 2). It will be shown that within a scan the change in OPD of the measurement interferometer can be determined to a resolution of 40 nm.
15
GUIDE TO THE EXPRESSION OF UNCERTAINTY IN MEASUREMENT
Saudi Standard · 2006 · 6.5K citations
Refractive index of air: new equations for the visible and near infrared
P.E. Ciddor · Applied Optics · 1996 · 1.4K citations