Abstract

We designed and constructed a single-fiber-optic confocal microscope (SFCM) with a microelectromechanical system (MEMS) scanner and a miniature objective lens. Axial and lateral resolution values for the system were experimentally measured to be 9.55 mum and 0.83 mum respectively, in good agreement with theoretical predictions. Reflectance images were acquired at a rate of 8 frames per second, over a 140 mum x 70 mum field-of-view. In anticipation of future applications in oral cancer detection, we imaged ex vivo and in vivo human oral tissue with the SFCM, demonstrating the ability of the system to resolve cellular detail.