Review of Scientific Instruments · 2002 · 43 citations · 31 references
Natural SciencesMaterials ScienceOptoelectronicsPhotonicsSpectroscopyInfrared SpectroscopyNear-infrared SpectroscopyOptical SpectroscopySemiconductor PhysicsTime-resolved SpectroscopyUltrafast OpticsCarrier Recombination DynamicsElectron-hole RecombinationFull Width At Half MaximumLaser Pulse DurationPump-probe SpectroscopyTemporal Resolution LimitsTime-resolved Infrared SpectroscopyUltrafast Spectroscopy
A facility for subnanosecond time-resolved (pump-probe) infrared spectroscopy has been developed at the National Synchrotron Light Source of Brookhaven National Laboratory. A mode-locked Ti:sapphire laser produces 2 ps duration, tunable near-infrared pump pulses synchronized to probe pulses from a synchrotron storage ring. The facility is unique on account of the broadband infrared from the synchrotron, which allows the entire spectral range from 2 cm−1 (0.25 meV) to 20 000 cm−1 (2.5 eV) to be probed. A temporal resolution of 100 ps, limited by the infrared synchrotron-pulse duration (full width at half maximum), is achievable. A maximum time delay of 170 ns is available without gating the infrared detector. To illustrate the performance of the facility, a measurement of electron–hole recombination dynamics for an HgCdTe semiconductor film in the far- and mid-infrared range is presented.
31
F. Wooten, Sumner P. Davis · American Journal of Physics · 1973 · 4.6K citations
B. S. Wherrett · Optica Acta International Journal of Optics · 1973 · 250 citations