In vivo depth-resolved oxygen saturation by dual-wavelength photothermal (DWP) OCT

Roman V. Kuranov, Shams Kazmi, Austin McElroy, Jeffrey W. Kiel, Andrew K. Dunn, Thomas E. Milner, Timothy Q. Duong

Optics Express · 2011 · 26 citations · 46 references

DOIFull text

Open access

Concepts

Abstract

Microvasculature hemoglobin oxygen saturation (SaO2) is important in the progression of various pathologies. Non-invasive depth-resolved measurement of SaO2 levels in tissue microvasculature has the potential to provide early biomarkers and a better understanding of the pathophysiological processes allowing improved diagnostics and prediction of disease progression. We report proof-of-concept in vivo depth-resolved measurement of SaO(2) levels in selected 30 µm diameter arterioles in the murine brain using Dual-Wavelength Photothermal (DWP) Optical Coherence Tomography (OCT) with 800 nm and 770 nm photothermal excitation wavelengths. Depth location of back-reflected light from a target arteriole was confirmed using Doppler and speckle contrast OCT images. SaO(2) measured in a murine arteriole with DWP-OCT is linearly correlated (R(2)=0.98) with systemic SaO(2) values recorded by a pulse-oximeter. DWP-OCT are steadily lower (10.1%) than systemic SaO(2) values except during pure oxygen breathing. DWP-OCT is insensitive to OCT intensity variations and is a candidate approach for in vivo depth-resolved quantitative imaging of microvascular SaO(2) levels.

References

46