Frequency of Goldmann Applanation Tonometer Calibration Error Checks
37
Citations
9
References
2005
Year
Goldmann applanation tonometers are not as accurate as the manufacturer's recommended calibration error tolerance of +/-0.5 mm Hg would suggest. Calibration error of less than +/-2.5 mm Hg is clinically acceptable. Calibration error checks should be carried out once monthly and tonometers with calibration error greater than +/-2.5 mm Hg returned to the manufacturer for re-calibration. Additional checks should be made if tonometers suffer specific damage. Ideally individual ophthalmologists should check calibration before each session.