Simultaneous calibration of optical tweezers spring constant and position detector response
24
Citations
10
References
2010
Year
Sensor CalibrationPosition Detector ResponseImpulse ResponseSingle LaserEngineeringMeasurementCalibrationMicroscopySimultaneous CalibrationEducationOptical TrappingLaser-based SensorInstrumentationOptical SystemsParticle Position DetectionOptical SensorOptical Sensors
We demonstrate a fast and direct calibration method for systems using a single laser for optical tweezers and particle position detection. The method takes direct advantage of back-focal-plane interferometry measuring not an absolute but a differential position, i.e. the position of the trapped particle relative to the center of the optical tweezers. Therefore, a fast step-wise motion of the optical tweezers yields the impulse response of the trapped particle. Calibration parameters such as the detector's spatial and temporal response and the spring constant of the optical tweezers then follow readily from fitting the measured impulse response.