Specific volume measurements of Cu, Mo, Pd, and Ag and calibration of the ruby <i>R</i>1 fluorescence pressure gauge from 0.06 to 1 Mbar
2.2K
Citations
6
References
1978
Year
Optical MaterialsIsothermal EquationsEngineeringMeasurementAbsorption SpectroscopySpectroscopic PropertySpecific Volume MeasurementsAnalytical InstrumentationCalibrationOptical DiagnosticsOptical PropertiesDiamond-window Pressure CellInstrumentationOptical SpectroscopyBiophysicsFluorescence Pressure GaugePhysicsThermal PhysicsNatural SciencesSpectroscopyApplied Physics
The wavelength shift with pressure of the ruby R1 fluorescence line (Δλ) has been calibrated in the diamond-window pressure cell from 0.06 to 1 Mbar. This was done by simultaneously making specific volume measurements of four metals (Cu, Mo, Ag, and Pd) and referring these results to isothermal equations of state derived from shock-wave experiments. The result is P (Mbar) = (19.04/5) {[(λ0+Δλ)/λ0]5−1}, where λ0 is the wavelength measured at 1 bar.