Metrology and Measurement Systems · 2014 · 21 citations · 7 references
Sensor CalibrationNon-uniformity CorrectionEngineeringSensorsMeasurementCalibrationTemperature MeasurementEducationThermal PhysicsRadiometryInstrumentationOutput SignalTemperature DriftThermal SensorDetector Array
Abstract In the article a non-uniformity correction method is presented which allows to compensate for the influence of detector’s temperature drift. For this purpose, dependency between output signal value and the temperature of the detector array was investigated. Additionally the influence of the temperature on the Offset and Gain coefficients was measured. Presented method utilizes estimated dependency between output signal of detectors and their temperature. In the presented method, the shutter is used for establishing signal reference. Thermoelectric cooler is used for changing the temperature of the detector array.
7
Scene-based nonuniformity correction method using multiscale constant statistics
Chao Zuo, Qian Chen, Guohua Gu et al. · Optical Engineering · 2011 · 32 citations