Optimization of confocal epifluorescence microscopy for microchip-based miniaturized total analysis systems
132
Citations
29
References
1998
Year
EngineeringMicroscopyAnalytical MicrosystemsMicro Total Analysis SystemLaser ApplicationsOptical MetrologyPinhole DiameterBiomedical EngineeringFiber OpticsMicroscopy MethodOptical DiagnosticsBiomedical OpticOptical SystemsLight MicroscopyMicrofluidicsBiophysicsInjection PlugApodized AperturesLaser MicroscopyBiophotonicsOptical ComponentsOptical SensorsOptical ImagingFluorescence MicroscopyConfocal Epifluorescence MicroscopyMicroscope Image ProcessingBiomedical ImagingBiomedical PhotonicsPlanar Glass SubstratesMedicine
The study presents a confocal epifluorescence detection scheme optimized for sub‑picomolar sensitivity in 30 µm‑deep planar glass microchips. The system uses a ×40, 0.6 NA objective under‑filled to an effective NA 0.22, a 12 µm laser spot, pinholes ranging 100–600 µm yielding probe volumes 0.1–2.4 pl, and quick‑fit flangeless fittings for interfacing tubing and capillaries to the glass channels. In continuous‑flow mode with a 400 µm pinhole, fluorescence intensity correlated linearly with fluorescein concentration (R² = 0.996, slope 0.85), achieving a 300 fM detection limit (34.6 ± 8.1 mV, S/N 6.1) and a 1 pM capillary zone electrophoresis S/N 5.8, with an injection plug of ~5470 molecules yielding ~570 detected molecules.
A confocal epifluorescence detection scheme, optimized to give sub-picomolar detection within planar glass substrates etched to a 30 µm depth, is described. A ×40, 0.6 numerical aperture (N.A.) lens with a 3.7 mm working distance was used to create a focused laser spot about 12 µm in diameter, by under-filling the lens aperture to give an effective, measured N.A. of 0.22 for the laser beam. The sectioning power (optical axis field of view) of various pinholes and the corresponding detector probe volumes (overlap of the excitation and observation volumes) were: (pinhole diameter, sectioning power, probe volume): 100 µm, 18 µm, 0.1 pl; 200 µm, 20 µm, 0.4 pl; 400 µm, 26 µm, 1.7 pl; and 600 µm, 36 µm, 2.4 pl. A log–log plot of fluorescence intensity versus fluorescein concentration, measured in continuous-flow mode using the optimum 400 µm pinhole, showed a correlation coefficient of 0.996 and a slope of 0.85. In this mode, 300 fM fluorescein gave a signal of 34.6 ± 8.1 mV over background with an S/N of 6.1, representing the lowest measured fluorescein dye concentration reported on-chip. Capillary zone electrophoresis of 1 pM fluorescein resulted in a mean S/N of 5.8. The injection plug, estimated to be about 5470 molecules, corresponds to 570 detected molecules on average. The design and use of quick-fit, flangeless fittings for interfacing tubing, fused-silica capillaries or pressurized systems to microfluidic channels etched in planar glass chips is briefly presented.
| Year | Citations | |
|---|---|---|
Page 1
Page 1