Sensitive Detection of Near-Infrared Fluorescent Dyes Using High-Performance Liquid Chromatography with Peroxyoxalate Chemiluminescence Detection System

Kohsuke Kimoto, Ryoya Gohda, Katsuhisa Murayama, Tomofumi Santa, Takeshi Fukushima, Hiroshi Homma, Kazuhiro Imai

Biomedical Chromatography · 1996 · 12 citations · 5 references

Concepts

Abstract

The highly sensitive detection of four near-infrared (near-IR) fluorescent dyes, methylene blue, pyridine 1, oxazine 1 and 3,3'-diethylthiadicarbocyanine iodide (DTDCI) by high-performance liquid chromatography (HPLC) with post-column peroxyoxalate chemiluminescence (PO-CL) detection utilizing bis(4-nitro-2-(3, 6,9-trioxadecyloxycarbonyl)phenyl) oxalate (TDPO) and hydrogen peroxide was examined. The detection limits for methylene blue, pyridine 1, oxazine 1 and DTDCI were 120, 27, 31, 0.19 fmol on-column at a signal-to-noise ratio of 2, respectively. The sensitivity for DTDCI was 250 times that obtained by HPLC with the conventional fluorescent detection.