91
Citations
8
References
2003
Year
The range of TSB was limited to 19.6 mg/dL and to 17 mg/dL for preterm infants. The correlation coefficients for all subjects (r = 0.94) and for term and preterm subjects between TcB measured by JM-103 and TSB was higher than that between TcB measured by JM-102 and TSB. The regression line in term infants between TcB measured by JM-103 and TcB was similar to that in preterm infants. The error distribution of TcB measured by JM-102 and TSB for all subjects (0.00 +/- 2.21) and for term and preterm subjects was larger than that of TcB measured by JM-103 and TSB (all subjects, 0.30 +/- 1.55).