Magnetic Resonance in Medicine · 2014 · 48 citations · 30 references
MedicineNeurobiologyNeuroscienceBiomedical ImagingDiagnostic ImagingSpectroscopic AnalysisBrain ImagingMagnetic Resonance ImagingMagnetic Resonance SpectroscopyEcho Planar ImagingFeasibility Study DesignIntraclass CorrelationIntrasubject Variability EstimationMagnetic Resonance Spectroscopic ImagingMetabolic Brain ImagingMetabolite ImagingSpectroscopic Imaging
PURPOSE: A feasibility study of an echo-planar spectroscopic imaging (EPSI) using a short echo time (TE) that trades off sensitivity, compared with other short-TE methods, to achieve whole brain coverage using inversion recovery and spatial oversampling to control lipid bleeding. METHODS: Twenty subjects were scanned to examine intersubject variance. One subject was scanned five times to examine intrasubject reproducibility. Data were analyzed to determine coefficients of variance (COV) and intraclass correlation coefficient (ICC) for N-acetylaspartate (NAA), total creatine (tCr), total choline (tCho), glutamine/glutamate (Glx), and myo-inositol (mI). Regional metabolite concentrations were derived by using multi-voxel analysis based on lobar-level anatomic regions. RESULTS: For whole-brain mean values, the intrasubject COVs were 14%, 15%, and 20% for NAA, tCr, and tCho, respectively, and 31% for Glx and mI. The intersubject COVs were up to 6% higher. For regional distributions, the intrasubject COVs were ≤ 5% for NAA, tCr, and tCho; ≤ 9% for Glx; and ≤15% for mI, with about 6% higher intersubject COVs. The ICCs of 5 metabolites were ≥ 0.7, indicating the reliability of the measurements. CONCLUSION: The present EPSI method enables estimation of the whole-brain metabolite distributions, including Glx and mI with small voxel size, and a reasonable scan time and reproducibility.
30
Design and construction of a realistic digital brain phantom
D. Louis Collins, Alex Zijdenbos, Vasken Kollokian et al. · IEEE Transactions on Medical Imaging · 1998 · 1.8K citations
Spatial mapping of the chemical shift in NMR
P. Mansfield · Magnetic Resonance in Medicine · 1984 · 398 citations