The American Journal of Sports Medicine · 1999 · 318 citations · 17 references
MedicineOrthopaedic SurgeryOrthopedic SurgeryBiomechanicsOrthopedic BiomechanicsOrthopedicsFixation BiomechanicsTendon MechanicsTendon RepairTendon SurgeryAnterior Cruciate Ligament TestsFixation StiffnessGraft TensioningScrew FixationTendon Graft FixationTibial Component Fixation
This study compared the stiffness (K), yield load (YL), and slippage (SL) of six tibial fixation methods. These properties were determined from load-to-failure and cyclic tests of double-looped tendon grafts fixed to both animal and young human tissue. Tandem washers (K = 259 N/mm, YL = 1159 N, SL = 0.5 mm) and the Washerloc (K = 248 N/mm, YL = 905 N, SL = 2.0 mm) were the two best fixations. At 500 N of load, which is the estimated daily tension of an anterior cruciate ligament graft during intensive rehabilitation, slippage was significantly greater in either of the other two methods for sutures tied to a post (4.9 mm), double staples (3.3 mm), and a 20-mm spiked metal washer (3.5 mm). Interference screw fixation performed well in animal tissue (YL = 776 N), but was significantly worse in young human tissue (YL = 350 N), with 57% of the fixations failing before 500 N of load. Animal tissue should not be used to estimate the performance of interference screw fixation in human tissue. Because 57% of the interference screw fixations using human tissue failed at loads below 500 N, their ability to provide adequate fixation during intensive rehabilitation should be questioned. However, both the Washerloc and tandem washers and screws provide fixation structural properties in young human tibia that should be appropriate for intensive rehabilitation.
17
Anterior Cruciate Ligament Graft Fixation
Mark E. Steiner, Aaron T. Hecker, Charles H. Brown et al. · The American Journal of Sports Medicine · 1994 · 398 citations
Yasuyuki Ishibashi, T. Rudy, Glen A. Livesay et al. · Arthroscopy The Journal of Arthroscopic and Related Surgery · 1997 · 286 citations