Current Protocols in Cytometry · 2005 · 83 citations · 51 references
EngineeringCell BiologyBiomedical OpticsInstrumentationMetrologyCalibrationFlow CytometryFlow MeasurementFlow MetrologyMeasurement StandardsParticle SizeFlow Cytometric Cell AnalysisFluorescence CalibrationOptical AlignmentOptical Axis AlignmentParticle Size CalibrationParticle Size MeasurementParticle Size Standards
Because flow cytometers are designed to measure particle characteristics, particles are the most common materials used to calibrate, control, and standardize the instruments. Definitions are provided for common terms to alert the reader to critical distinctions in meaning. The unit presents extensive background on particle types and cautions and goes on to describe practical aspects of methods to standardize and calibrate instruments, in terms of optical alignment, fluorescence and light scatter resolution, and sensitivity. Finally, suggestions follow for analyzing particles used for calibration.
51
Caroline Dive · British Journal of Cancer · 1991 · 671 citations · Full text
F. Borek · Journal of Immunological Methods · 1980 · 231 citations