Concepedia

Publication | Closed Access

Carbon Rod Atomizer Applied to Measurement of Lead in Whole Blood by Atomic Absorption Spectrophotometry

46

Citations

2

References

1972

Year

Abstract

Abstract An atomic absorption method is described for determining lead in blood by means of the carbon rod atomizer. With the procedure, only a dilution of the whole blood sample is required, and results are comparable to those obtained by the more generally used atomic absorption flame technique. Advantages of the carbon rod include rapid analysis, simple sample preparation, and small sample volumes.