The Journal of Maternal-Fetal & Neonatal Medicine · 2015 · 16 citations · 7 references
MedicineChild HealthPediatricsNeonatal HealthNeonatologyResuscitationCardiopulmonary Exercise TestingCardiopulmonary ResuscitationNeonatal ResuscitationResuscitation MonitoringForce Application PointMechanical Cardiopulmonary ResuscitationMechanical Chest Compression DevicesNeonatal Cardiopulmonary ResuscitationNeonatal Resuscitation SafetyNeonatal Resuscitation SupportPediatric Cardiopulmonary ResuscitationPost-resuscitation Physiologic Monitoring
OBJECTIVE: To assess utility of the FingerTPS™ system in measuring chest compression (CC) rate and force. METHODS: Five minutes of CC was performed in a neonatal manikin without (n = 29) and with (n = 30) a metronome. The FingerTPS™ force (lbs.) was compared to pressure (mmHg) in a 50-mL normal-saline bag inside the manikin. FingerTPS™ CC rate and the time until a 20% decline from baseline force and pressure were calculated. RESULTS: The normal-saline pressure declined earlier than the FingerTPS™ force. Metronome use did not influence CC rate, force or pressure. CONCLUSIONS: The FingerTPS™ can be used to measure CC rate and force.
7
Catherine Christman, Rae Jean Hemway, Myra H. Wyckoff et al. · Archives of Disease in Childhood Fetal & Neonatal · 2010 · 96 citations
Rescuer fatigue during simulated neonatal cardiopulmonary resuscitation
Elliott Li, Michael O’Reilly, Khalid Aziz et al. · Journal of Perinatology · 2014 · 44 citations
Charles Christoph Roehr, Georg M. Schmölzer, Marta Thió et al. · Journal of Paediatrics and Child Health · 2014 · 26 citations