Quantifying force application to a newborn manikin during simulated cardiopulmonary resuscitation

Anne Lee Solevåg, Po‐Yin Cheung, Elliott Li, Khalid Aziz, Megan O’Reilly, Bo Fu, Bin Zheng, Georg M. Schmölzer

The Journal of Maternal-Fetal & Neonatal Medicine · 2015 · 16 citations · 7 references

Concepts

Abstract

OBJECTIVE: To assess utility of the FingerTPS™ system in measuring chest compression (CC) rate and force. METHODS: Five minutes of CC was performed in a neonatal manikin without (n = 29) and with (n = 30) a metronome. The FingerTPS™ force (lbs.) was compared to pressure (mmHg) in a 50-mL normal-saline bag inside the manikin. FingerTPS™ CC rate and the time until a 20% decline from baseline force and pressure were calculated. RESULTS: The normal-saline pressure declined earlier than the FingerTPS™ force. Metronome use did not influence CC rate, force or pressure. CONCLUSIONS: The FingerTPS™ can be used to measure CC rate and force.