Sequential Voltammetric Determination of Uranium, Cadmium and Lead by Using the <i>ex situ</i> Bismuth Film Electrode: Application to Phosphate Fertilizers

Cristiane Luisa Jost, Lara M. di Martos, Luis Ferraz, Paulo Cı́cero do Nascimento

Electroanalysis · 2015 · 16 citations · 47 references

Concepts

Abstract

Abstract A sequential voltammetric procedure for the determination of uranium, cadmium and lead was investigated at an ex situ bismuth film electrode (BiFE). First, the adsorptive stripping voltammetry was applied to assay the U(VI)‐cupferron complex in the differential pulse mode (detection limit of 1.0 µg L −1 , 200 s accumulation time). Through the manipulation of the same aliquot of the sample, efforts were made to quantify cadmium and lead by square wave anodic stripping voltammetry. Detection limits of 2.03 µg L −1 for Cd (II) and 2.43 µg L −1 for Pb (II) were calculated (100 s accumulation time). The methodology was successfully applied to phosphate fertilizer samples after open vessel wet decomposition (HNO 3 /H 2 O 2 ). The following value ranges were evaluated: U (VI) 37.2–150 mg kg −1 , Pb (II) 78.3–204 mg kg −1 and Cd (II) 44.1–71.6 mg kg −1 . Validation was performed by using the standard reference materials SRM‐695 – phosphate fertilizer – and SRM‐1643e – water.

References

47