Cardiac imaging at 7 tesla: Single‐ and two‐spoke radiofrequency pulse design with 16‐channel parallel excitation

Sebastian Schmitter, Lance DelaBarre, Xiaoping Wu, Andreas Greiser, Dingxin Wang, Edward J. Auerbach, J. Thomas Vaughan, Kǎmil Uǧurbil, Pierre‐François Van de Moortele

Magnetic Resonance in Medicine · 2013 · 71 citations · 32 references

DOIFull text

Open access

Concepts

Abstract

PURPOSE: Higher signal to noise ratio (SNR) and improved contrast have been demonstrated at ultra-high magnetic fields (≥7 Tesla [T]) in multiple targets, often with multi-channel transmit methods to address the deleterious impact on tissue contrast due to spatial variations in B1 (+) profiles. When imaging the heart at 7T, however, respiratory and cardiac motion, as well as B0 inhomogeneity, greatly increase the methodological challenge. In this study we compare two-spoke parallel transmit (pTX) RF pulses with static B1 (+) shimming in cardiac imaging at 7T. METHODS: Using a 16-channel pTX system, slice-selective two-spoke pTX pulses and static B1 (+) shimming were applied in cardiac CINE imaging. B1 (+) and B0 mapping required modified cardiac triggered sequences. Excitation homogeneity and RF energy were compared in different imaging orientations. RESULTS: Two-spoke pulses provide higher excitation homogeneity than B1 (+) shimming, especially in the more challenging posterior region of the heart. The peak value of channel-wise RF energy was reduced, allowing for a higher flip angle, hence increased tissue contrast. Image quality with two-spoke excitation proved to be stable throughout the entire cardiac cycle. CONCLUSION: Two-spoke pTX excitation has been successfully demonstrated in the human heart at 7T, with improved image quality and reduced RF pulse energy when compared with B1 (+) shimming.

References

32