2002 · 18 citations · 4 references
EngineeringBiomedical EngineeringElectrical EngineeringBiomedical ImagingDiagnostic ImagingPower ElectronicsConverter TopologiesDc-dc ConversionDc-dc ConvertersMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsVoltage RegulatorsBuck ConvertersFull-bridge ConvertersHigh-efficiency Power AmplifiersMagnetic ResonanceMagnetic Resonance Imaging InstrumentationMagnetic Resonance PhysicsSwitching LossesSwitching-induced Losses
A high-power, high-efficiency switching amplifier capable of generating arbitrary gradient coil waveforms for a magnetic resonance imaging system is described. A high-frequency "partial-voltage" buck regulator feeds a "full-voltage" full-bridge inverter to relieve the full bridge of the necessity of switching when high voltage is needed by the load. When high voltage is needed, the partial-voltage buck circuit that employs faster lower voltage IGBT devices assumes the switching duty. When low voltage is needed, the high-voltage bridge inverter, operating from a low-voltage DC bus, assumes the switching duty. Since switching losses are approximately proportional to bridge input voltage, the switching losses incurred by the inverter are minimized.
4
Limits to neural stimulation in echo‐planar imaging
P. Mansfield, Paul R. Harvey · Magnetic Resonance in Medicine · 1993 · 100 citations
Quasi-linear IGBT inverter topologies
O. Mueller, J.N. Park · 2002 · 64 citations
Engineering, Biomedical Engineering, Electrical Engineering +16
A GTO speed-up inverter for fast-scan magnetic resonance imaging
O. Mueller, Peter Roemer, J.N. Park et al. · 2003 · 14 citations