EngineeringBiomedical EngineeringElectrical EngineeringBiomedical ImagingDiagnostic ImagingPower ElectronicsConverter TopologiesDc-ac InvertersMagnetic Resonance ImagingMedical ImagingPower AmplifiersPulse-width ModulationHigh-efficiency Power AmplifiersHigh-frequency ConversionInverter SystemsMagnificationPulse Sequence DesignSinusoidal Pulse-width ModulationVariable-frequency Pulse-width Modulation
Magnetic resonance imaging (MRI) systems require stronger gradient coil currents with faster ramp times for high-speed image taking and in order to obtain a smaller field of view or a magnification effect. Previous speedup inverter circuits able to provide those higher power levels have been discussed at APEC-92 and APEC-93. They could generate squarewave currents with a constant ramp slope determined by the applied high voltage and the coil inductance. Two new circuit topologies are described here which permit the generation of variable ramp as well as sinusoidal current waveforms for the benefit of pulse sequence designers. The result is a novel high-efficiency quasi-linear power amplifier concept. In one case high-frequency pulse-width modulation schemes are replaced by low-frequency baseband switching. Very good sinusoidal current waveforms (150 A peak, 1150 V) have been obtained.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
4
Limits to neural stimulation in echo‐planar imaging
P. Mansfield, Paul R. Harvey · Magnetic Resonance in Medicine · 1993 · 100 citations
A GTO speed-up inverter for fast-scan magnetic resonance imaging
O. Mueller, Peter Roemer, J.N. Park et al. · 2003 · 14 citations