Journal of Clinical Microbiology · 1984 · 17 citations · 17 references
Veterinary MedicineInfectious DiseasesMicrobiologyDiagnostic MicrobiologyMedical MicrobiologyVirologyAntigen Detection AssaysComplement SystemDiagnostic ImmunologyViral DiagnosticsViral InfectionsVirus PurificationComplement Fixation TestsDiagnostic Test ProceduresHemagglutination AssaysHemagglutination Inhibition AssaysNeutralizationNeutralizing Antibody AssaysNeutralizing Antibody MeasurementViral Antigen Detection
We describe here a microneutralization procedure for conveniently testing large numbers of specimens for antibodies to enterovirus 70. The test utilized human rhabdomyosarcoma cells and was read by staining with crystal violet after 4 days of incubation. The test compares well with other serological assays, being more sensitive than the standard tube neutralization test and the complement fixation test, but less sensitive than the hemagglutination inhibition test. However, the hemagglutination inhibition test required concentrated, partially purified virus as antigen, as did the complement fixation test, and was difficult to read, so that its greater sensitivity may not be of practical significance. By all four test procedures, a recent isolate of enterovirus 70 was a more sensitive antigen than the prototype strain, as shown by greater geometric mean titers in sera of patients from various epidemics.
17
Enterovirus type 70: the etiologic agent of pandemic acute haemorrhagic conjunctivitis.
Radmila R. Mirkovic, Reisaku Kono, M Yin-Murphy et al. · PubMed · 1973 · 103 citations
PANDEMIC OF NEW TYPE OF CONJUNCTIVITIS
Reisaku Kono, Akira Sasagawa, K. Ishii et al. · The Lancet · 1972 · 98 citations