Estimation of oxyhaemoglobin and of methaemoglobin by a photoelectric method

B. Szigeti

Biochemical Journal · 1940 · 24 citations · 0 references

DOIFull text

Open access

Concepts

Abstract

THE oxyhaemoglobin content of blood can be rapidly and accurately estimated with the help of a photoelectric colorimeter. In the absence of oxyhaemoglobin or of other pigments, methaemoglobin can be determined in the same way. In whole blood the presence of met-or of sulph-haemoglobin or of other pigments can be detected. The experiments also indicate that the use of 084 % ammonia as a diluent for blood for spectrophotometric measurements is not safe. Better results can be obtained with a concentration of 0 04 % ammonia.