Enzymatic Immunoassay vs. Gas/Liquid Chromatography for Determination of Phenobarbital and Diphenylhydantoin in Serum

Harold E. Booker, Barbara A. Darcey

Clinical Chemistry · 1975 · 52 citations · 0 references

DOIFull text

Open access

Concepts

Abstract

The antiepileptic drugs diphenylhydantoin and phenobarbital were measured in serum by a new commercially available enzyme immunoassay procedure ("EMIT," Syva Corp.). The procedure requires less than 5 min and no more than 50 mul of serum per determination. It is simple; only four steps (pipetting and diluting with an automatic pipettor-dilutor) are required before spectrophotometry. Twenty replicate analyses of a serum containing phenobarbital and diphenylhydantoin gave results with a CV of 6.8% and 9.1%, respectively. Results attained in a large series of patients were compared with results by a gas/liquid chromatographic procedure. For phenobarbital r = 0.97, and for diphenylhydantoin r = 0.98. No false negatives or false positives were encountered.