Preparation of Graphite Targets from Small Marine Samples for AMS Radiocarbon Measurements

L. Liong Wee Kwong, Pavel P. Povinec, J A Timothy Jull

Radiocarbon · 2004 · 21 citations · 17 references

DOIFull text

Open access

Concepts

Abstract

A vacuum sample processing line was set up and methods were developed for the determination of radiocarbon in small-volume seawater and biota samples. Seawater samples (500 mL per borosilicate glass bottle and poisoned with HgCl 2 ) were acidified with 5 mL concentrated hydrochloric acid. Pure N 2 was used as a carrier gas to strip CO 2 from the samples for 10 min in a circulation mode. After purification through several water traps, the CO 2 was isolated cryogenically. Using Na 2 CO 3 standard solutions, recovery yields were calculated superior to 95 ± 5%. Freeze-dried marine biota samples were thoroughly mixed with Cu(II)O and combusted at 900°. The CO 2 was purified by passing through Ag wool and Cu granules at 450° before reduction to graphite. Finally, graphite was synthesized using Zn dust heated to 450° in the presence of an Fe catalyst at 550°. Although this method takes about 8 hr (synthesis done overnight), the advantage is that no water vapor by-product is formed to hinder the reaction. The graphite yields, measured both by gravimetric methods and by pressure readings, were 95 ± 5%. Accelerator mass spectrometry (AMS) measurements were carried out at the NSF-Arizona AMS Facility. Results for water samples from the northwest Pacific Ocean are reported which are in agreement with data reported elsewhere.

References

17