2004 · 19 citations · 9 references
EngineeringBiomedical EngineeringElectrical EngineeringBioinstrumentationMotor NeuroscienceNeural EngineeringFunctional Electrical StimulationImplantable Biomedical DevicesImplantable Neural DevicesNeural ProstheticsNeural StimulationNeurostimulationCharge Pump CircuitsCharge Pump DesignImplantable Functional Electrical StimulationImplantable Neural ElectrodesImplantable Neural StimulationNeural Stimulation DevicesSafety ConstraintsSafety Of Implantable Stimulation
We propose a new implantable circuit for the internal functional electrical stimulation (FES) of motor nerves for paraplegic people. The circuit is designed to deliver precise calibrated stimulation pulses to specific multipolar electrodes. Several original design features have been developed to respond to the particular specifications imposed by safety constraints. In particular, the DAC has been thought to be fully monotonic and the output stage to ensure a passive and secure discharge of the safety capacitor. Also some features have been added in order to improve the classical charge pump that generates on-chip high voltage.
9
BION™ system for distributed neural prosthetic interfaces
Gerald E. Loeb, Raymond A. Peck, William H. Moore et al. · Medical Engineering & Physics · 2001 · 245 citations