Antarctica A Keystone in a Changing World · 1996 · 74 citations · 1 references
Natural SciencesChemistryAnalytical ChemistryElemental AnalysisWater TreatmentEnvironmental Analytical ChemistryHeavy Metal AnalysisSample HandlingSample PreparationTrace Element AnalysisWater QualityAcid DigestionAnalytical Detection Of Heavy MetalsDigestionEnvironmental Sample PreparationSample Contamination AssessmentTrace Metal ContaminationWater Quality Sampling
Water samples for trace-metal determinations routinely have been prepared in open laboratories. For example, the U.S. Geological Survey method 1-3485-85 (Extraction Procedure, for Water-Suspended Sediment) is performed in a laboratory hood on a laboratory bench without any special precautions to control airborne contamination. This method tends to be contamination prone for several trace metals primarily because the samples are transferred, acidified, digested, and filtered in an open laboratory environment. To reduce trace-metal contamination of digested water samples, procedures were established that rely on minimizing sample-transfer steps and using a class-100 clean bench during sample filtration. This new procedure involves the following steps:
1