Publication | Closed Access
A Glucose Meter Accuracy and Precision Comparison: The FreeStyle Flash Versus the Accu-Chek Advantage, Accu-Chek Compact Plus, Ascensia Contour, and the BD Logic
49
Citations
19
References
2008
Year
No statistically significant differences in accuracy were evident using the Clarke EGA criteria (pooled results of Zone A and B), though the more strict 20% accuracy criteria (Zone A results only) found the Flash and Contour to have significantly greater accuracy compared to the Advantage, Compact Plus, and the Logic. Using the 10% accuracy criteria found the Flash to have significantly greater accuracy than each of the other four meters. All five meters demonstrated similar precision.
| Year | Citations | |
|---|---|---|
Page 1