PubMed · 2007 · 10 citations · 8 references
PharmacyChemistryPharmaceutical AnalysisAnalytical ChemistryChromatographic AnalysisImpurity ProfilingAnalytical ChromatographyChromatographic Data AnalysisChromatographic Impurity SeparationChromatographic Method ValidationChromatographic PrinciplesChromatographyAssayChromatographic BehaviorChromatographic ValidationHigh-performance Liquid ChromatographyHigh-performance Liquid Chromatography Method DevelopmentImpuritiesIn-process Quality ControlIsocratic Method Validation
A simple, rapid and specific ion-pair HPLC method for the determination of (R,R)-glycopyrronium bromide and its related impurities is presented, and parameters affecting the chromatographic properties of these compounds are discussed. Optimal analyte separation was achieved on base deactivated Nucleosil at 40 degrees C, using phosphate buffer pH 2.30 with sodium-1-decanesulfonate (0.01 M)/methanol (35/65; v/v) as eluent for isocratic elution at a flow rate 1 ml x min(-1). The analytical assay was validated according to international guidelines. The methodis suitable for in-process control and as stability indicating assay.