Publication | Closed Access
Fabrication and testing of a large area, high density, parylene MEMS
44
Citations
6
References
2011
Year
Unknown Venue
Large AreaMedical ElectronicsEngineeringMechanical EngineeringMicroelectromechanical SystemsMaterial InnovationBiomedical EngineeringMedical InstrumentationSocial SciencesMicro-electromechanical SystemHigh DensityParylene MemsMicromachinesFlexible 256-Electrode ArrayBiomedical DevicesBio-electronic InterfacesMaterials ScienceElectrical EngineeringNanomanufacturingFabrication TechniqueµEcog GridNeural InterfaceMicro TechnologyBiomedical SensorsNeuroengineeringFlexible ElectronicsMicrofabricationBiomedical DiagnosticsBioelectronicsApplied PhysicsElectrode PitchElectrophysiologyBrain Electrophysiology
We report the design, microfabrication and testing of a flexible 256-electrode array for micro-electrocorticography (µECoG) with an electrode pitch of 500 µm. Our µECoG grid is a flexible five-layer parylene MEMS device (two layers of platinum insulated by three layers of parylene) featuring plasma-etched vias and a monolithically integrated parylene cable which is compression-bonded to a fan-out board using anisotropic conductive film (ACF) technology. We have characterized the device by electrochemical impedance spectroscopy in artificial cerebrospinal fluid (aCSF) and recorded acoustic evoked potentials in vivo from the rat primary auditory cortex.