ASAIO Journal · 2002 · 41 citations · 14 references
MedicineNephrologyDialysisHemodialysisRenal Replacement TherapyBlood Volume RegulationChronic HemodialysisIntradialytic HypotensionRenal Blood FlowRenal HemodynamicsUremia ManagementBiofeedbackCardiovascular Stability During HemodialysisDialysis-related HypotensionHemodynamic Response To UltrafiltrationHemodynamic Stability ManagementHypotension TreatmentIntradialytic Hypotension PreventionSystolic Blood Pressure
Intradialytic hypotension remains a frequent complication of dialysis, occurring in up to 33% of patients. We tested a fully integrated biofeedback system (the Hemocontrol system) that monitors and regulates blood volume contraction during hemodialysis. Seven hypotension prone patients were selected for the study. We conducted a prospective crossover study alternating dialysis sessions using the blood volume regulation system and standard dialysis sessions. Event free sessions were defined as dialysis sessions not requiring any therapeutic intervention for hypotension related signs or symptoms. There was a significant improvement in the number of event free sessions with blood volume regulation compared with standard dialysis (50.8% of sessions vs. 29.2%; p < 0.01). Percentages of event free sessions and mean postdialysis systolic blood pressure improved progressively over the course of the study, indicating improved hemodynamic stability over the study period. Therefore, the use of a biofeedback system to monitor and regulate blood volume during dialysis was helpful in restoring cardiovascular stability in a population of hypotension prone hemodialysis patients. Further studies are needed to confirm these preliminary results and to establish the role of blood volume regulation systems in reducing the incidence of hypotension during hemodialysis.
14
Hematocrit as an Indicator of Blood Volume and a Predictor of Intradialytic Morbid Events
Robert R. Steuer, John K. Leypoldt, Alfred K. Cheung et al. · ASAIO Journal · 1994 · 115 citations
Medicine, Nephrology, Dialysis +16
Blood Volume Regulation During Hemodialysis
Antonio Santoro, Francesco Paolini, Giovanni Cavicchioli et al. · American Journal of Kidney Diseases · 1998 · 112 citations
Medicine, Nephrology, Physiology +12
Hemoscan: a dialysis machine-integrated blood volume monitor.
F. Paolini, Antonio Bosetto, Antonio Santoro · PubMed · 1995 · 62 citations
Automatic Control of Blood Volume Trends during Hemodialysis
Antonio Santoro, Francesco Paolini, M. Spongano et al. · ASAIO Journal · 1994 · 52 citations
Medicine, Nephrology, Physiology +17