Determination of the in-flight spectral and radiometric characteristics of the Airborne Visible/Infrared Imaging Spectrometer (AVIRIS)

29

Citations

0

References

1991

Year

Abstract

AVIRIS is a science research imaging spectrometer that measures radiance in 224 channels between 400 to 2450 nm in the electromagnetic spectrum. To determine the inflight spectral and radiometric characteristics of AVIRIS a validation and calibration experiment was performed. Five data sets were acquired over a calibration site on the homogeneous playa of Rogers Dry Lake, California, U.S. Surface reflectance, atmospheric optical depths, and atmospheric water vapor measurements were acquired concurrently with the overflights. These in situ measurements were used to constrain the LOWTRAN 7 radiative transfer code to predict the total spectral radiance incident at the AVIRIS aperture. These predicted radiances and the AVIRIS measured radiances were analyzed to validate the inflight characteristics. Inflight spectral channel positions and response functions over the AVIRIS spectral range were derived. Radiometric calibration coefficients were calculated for each channel as well as radiometric accuracy, intraflight stability, and noise equivalent delta radiance.