2015 · 59 citations · 12 references
EngineeringCardiologyBioinstrumentationBiomedical Signal ProcessingExercise PhysiologyCardiovascular MonitoringDigital Signal ProcessingSignal ProcessingTime-frequency AnalysisWearable BiosensorsWearable Health TechnologiesAdaptive FiltersHeart RateHeart Rate MonitoringMotion ArtifactsPhotoplethysmogram Signal ProcessingPhotoplethysmographyPhysiological Signal ProcessingPulse Oximeters
Accurate and reliable estimation of the heart rate using wearable devices, especially during physical exercise, must deal with noisy signals that contain motion artifacts. We present an approach that is based on photoplethysmographic (PPG) signals which are measured with two wrist-type pulse oximeters. The heart rate is related to intensity changes of the reflected light. Our proposed method suppresses the motion artifacts by adaptively estimating the transfer functions of each of the three-axis acceleration signals that produce the artifacts in the PPG signals. We combined the output of the six adaptive filters into a single enhanced time-frequency domain signal based on which we track the heart rate with a high accuracy. Our approach is real-time capable, computationally efficient and real data results for a benchmark data set illustrate the superior performance compared to a recently proposed approach.
12