Femoral stem fixation

George Shen

Journal of Bone and Joint Surgery - British Volume · 1998 · 108 citations · 0 references

Concepts

Abstract

The excellent long-term results for the first-generation Charnley stem may not apply to later versions with flanges. It seems possible that the early design functioned as a taper-slip system, as accepted in the Exeter prosthesis. Comparison with the requirements for the alternative composite-beam system for the femoral component shows considerable differences that have important implications. These include design, surface finish, cementing technique and the interpretation of radiological signs of loosening. A distinction should be made between the requirements for the successful use of the two different engineering systems.