Japanese Journal of Radiological Technology · 2004 · 10 citations · 1 references
MedicineRadiologyRadiation DosimetryRadiation ProtectionRadiation SafetyAbsorbed DoseHealth PhysicsPatient-specific DosimetryRadiation Dose MeasurementRadiation Protection PrinciplesRadiation Risk AssessmentEnergy Absorption CoefficientFlat-panel DetectorsMass Energy-absorption CoefficientPatient Dose AssessmentRadiation Dose AssessmentRadiation Dose EstimationX-ray DosimetryX-ray Exposure Assessment
The IEC requires interventional X‑ray systems to display air kerma and dose‑area product, but converting these values to patient entrance skin dose necessitates accounting for effective energy, the skin‑to‑air mass‑energy absorption ratio, backscatter, and the characteristics of automatic exposure control. This study aimed to determine how clinical parameters—tube voltage, current, pulse width, additional filter, and focal spot size—vary with patient body size to enable accurate calculation of entrance skin dose. We measured the effective X‑ray energy for patients under different parameter settings to derive a conversion factor from displayed air kerma to skin dose. The resulting conversion factor from air kerma in free air to entrance skin dose is approximately 1.4, providing a practical estimate for radiation protection.
The International Electrotechnical Commission has stipulated that medical X-ray equipment for interventional procedures must display radiation doses such as air kerma in free air at the interventional reference point and dose area product to establish radiation safety for patients (IEC 60601-2-43). However, it is necessary to estimate entrance skin dose for the patient from air kerma for an accurate risk assessment of radiation skin injury. To estimate entrance skin dose from displayed air kerma in free air at the interventional reference point, it is necessary to consider effective energy, the ratio of the mass-energy absorption coefficient for skin and air, and the backscatter factor. In addition, since automatic exposure control is installed in medical X-ray equipment with flat panel detectors, it is necessary to know the characteristics of control to estimate exposure dose. In order to calculate entrance skin dose under various conditions, we investigated clinical parameters such as tube voltage, tube current, pulse width, additional filter, and focal spot size, as functions of patient body size. We also measured the effective energy of X-ray exposure for the patient as a function of clinical parameter settings. We found that the conversion factor from air kerma in free air to entrance skin dose is about 1.4 for protection.
1
Douglas W. Fletcher, Donald L. Miller, Stephen Balter et al. · Journal of Vascular and Interventional Radiology · 2002 · 131 citations