Image quality at low tube voltage (70 kV) and sinogram-affirmed iterative reconstruction for computed tomography in infants with congenital heart disease

Motoo Nakagawa, Yoshiyuki Ozawa, Keita Sakurai, Masashi Shimohira, Kazuya Ohashi, Miki Asano, Sachiko Yamaguchi, Yuta Shibamoto

Pediatric Radiology · 2015 · 32 citations · 16 references

DOIFull text

Open access

Concepts

Abstract

BACKGROUND: Lower tube voltage has advantages for CT angiography, such as improved contrast OBJECTIVE: To evaluate the image quality of low-voltage (70 kV) CT for congenital heart disease and the ability of sinogram-affirmed iterative reconstruction to improve image quality. MATERIALS AND METHODS: Forty-six children with congenital heart disease (median age: 109 days) were examined using dual-source CT. Scans were performed at 80 kV and 70 kV in 21 and 25 children, respectively. A nonionic iodinated contrast medium (300 mg I/ml) was used for the 80-kV protocol. The contrast medium was diluted to 75% (225 mgI/mL) with saline for the 70-kV protocol. Image noise was measured in the two protocols for each group by extracting the standard deviations of a region of interest placed on the descending aorta. We then determined whether sinogram-affirmed iterative reconstruction reduced the image noise at 70 kV. RESULTS: There was more noise at 70 kV than at 80 kV (29 ± 12 vs 20 ± 4.8; P < 0.01). Sinogram-affirmed iterative reconstruction with grade 4 strength settings improved the noise (20 ± 5.9; P < 0.01) for the 70-kV group. CONCLUSION: Sinogram-affirmed iterative reconstruction improved the image quality of CT in congenital heart disease.