Ophthalmic surgery, lasers & imaging retina · 2007 · 13 citations · 11 references
MedicineOphthalmologyVeterinary OphthalmologyClinical OptometryOphthalmic SurgerySurgical InstrumentationIntraocular Pressure MeasurementOcular BiometryRabbit OphthalmologyAnterior Chamber DimensionsInterobserver Reliability AssessmentInterobserver ReproducibilityIntraocular Pressure AssessmentManometersTonometry Error Sources
<H4>BACKGROUND AND OBJECTIVES</H4> <P>To validate accuracy and reproducibility of the Perkins tonometer, pneumatonometer, and Tono-Pen XL (Medtronic Solan, Jacksonville, FL) in estimating intraocular pressure (IOP) in rabbits. </P> <H4>MATERIALS AND METHODS</H4> <P> IOP was increased from 5 to 50 mm Hg in 5-mm increments. Measurements were compared to readings of two digital manometers simultaneously measuring real IOP in the anterior chamber and vitreous cavity. Interobserver accuracy was evaluated using 4 eyes with the Perkins tonometer. </P> <H4>RESULTS</H4> <P> The Perkins tonometer and Tono-Pen XL underestimated IOP and were more accurate at pressures less than 30 mm Hg. No statistically significant difference was found between real IOP and Tono-Pen XL readings. The pneumatonometer overestimated pressures in the low ranges but was accurate at pressures greater than 40 mm Hg. The Tono-Pen XL had more variability than the Perkins tonometer and pneumatonometer at high IOP. </P> <H4>CONCLUSIONS</H4> <P> None of the tonometers are accurate or reproducible in estimating IOP in rabbits over the tested range. Pneumatonometry, although not very accurate, has the advantage of having acceptable variability. </P> <P>[<CITE>Ophthalmic Surg Lasers Imaging</CITE> 2007;38:43-49.] </P> <H4>AUTHORS</H4> <P>From the Ophthalmic Biophysics Center (ACA, EME, IN, MO, HY, FF, J-MP), Bascom Palmer Eye Institute, University of Miami Miller School of Medicine, Miami, Florida; the Biostatistic Center (ASV), Bascom Palmer Eye Institute, University of Miami Miller School of Medicine, Miami, Florida; the Department of Biomedical Engineering (J-MP), University of Miami College of Engineering, Miami, Florida; and the University of Liege (J-MP), Liege, Belgium. </P> <P>Accepted for publication September 18, 2006. </P> <P>Supported by Florida Lions Eye Bank; InnFocus LLC, NIH Center Grant No. P30-EY014801, Research to Prevent Blindness, Inc.; and the Henri and Flore Lesieur Foundation. </P> <P>Address correspondence to Jean-Marie Parel, PhD, Ophthalmic Biophysics Center, Bascom Palmer Eye Institute, University of Miami, Miller School of Medicine, 1638 NW 10th Avenue, Miami, FL 33136. </P>
11
A Rat Model of Chronic Pressure-induced Optic Nerve Damage
John C. Morrison, Charity G. Moore, LISA M.H. DEPPMEIER et al. · Experimental Eye Research · 1997 · 524 citations
New implant for drainage in glaucoma. Animal trial.
Anthony CB Molteno · British Journal of Ophthalmology · 1969 · 271 citations · Full text
Medicine, Ophthalmology, Surgery +12