Two juvenile size silicone testicular prostheses were analyzed following replacement by adult size devices after 72 and 99 months in situ. Scanning electron microscopy and electron microprobe studies were carried out on the explanted silicone prostheses and their attached tissue capsules. The material at the surface of the prostheses gave no evidence of biodeterioration. However, localized sites of calcification were observed. The adjacent tissue capsule consisted of two distinct concentric layers with different morphologies; the one adjacent to the prosthesis consisted of dense arrays of collagen fibres oriented parallel to the silicone surface and extending outward to a thickness of 100-120 microns. The outer layer was less dense, more disordered and was uneven in thickness. It contained numerous discrete particles of silicon-rich material. Technical and clinical implications of these results are discussed.