Publication | Closed Access
Results from a multicenter evaluation of the 4th generation Elecsys Troponin T assay.
64
Citations
0
References
2007
Year
Our data confirm an excellent analytical performance of the improved troponin T immunoassay. Most importantly, no systematic bias between cTnT results determined in serum and heparin plasma was observed from data obtained in 7 evaluation sites. The performance of the 4th generation Elecsys Troponin T assay is therefore comparable to other commercially available troponin immunoassays. Further studies are necessary to investigate the cause of poor comparability of cTnT results in rare individual serum to matched heparin plasma samples.