A Noise-Temperature Measurement System Using a Cryogenic Attenuator

Jorge E. Fernández

1998 · 65 citations · 3 references

Concepts

Abstract

This article describes a method to obtain accurate and repeatable input noisetemperature measurements of cryogenically cooled low-noise ampliflers using a commercial noise-flgure meter in conjunction with a cryogenic attenuator co-located with the amplifler under test. The calibration techniques used also are discussed. I. Introduction Accurate and rapid noise-temperature measurements are critical to the modeling and development of high electron mobility transistor (HEMT) cryogenically cooled ampliflers. Four difierent techniques have been used in the laboratory to make noise-temperature measurements of cryogenically cooled ampliflers, and each of these is described below. Following this, the cryogenic pad method, the method most often used by the Cryogenics{Electronics Front-End Equipment Group at JPL, is described. Although the cryogenic pad and noise-flgure meter are not in and of themselves new, combining them provides an elegant, simple, and quick solution to equivalent input noise-temperature and gain measurements.