PharmacyPharmacologyClinical PharmacologyDrug Approval ProcessesGood Laboratory PracticeBioanalytical Method ValidationConfirmatory TestingEndotoxin ControlEndotoxin AssaysEndotoxin DetectionEndotoxin TestingEndotoxin ValidationLimulus Amebocyte Lysate AssayLimulus Amebocyte Lysate TestPyrogen Detection AssaysRabbit Pyrogen Test
The draft guideline sets forth acceptable conditions for use of the Limulus Amebocyte Lysate (LAL) test. It also describes procedures for using this methodology as an end-product endotoxin test for human (including biological products) and animal injectable drugs, and medical devices. Manufacturers are allowed to switch from the rabbit pyrogen test to the LAL test, without preclearance from FDA, if they follow the guideline.