Ophthalmic surgery, lasers & imaging retina · 1994 · 16 citations · 4 references
MedicineOphthalmologySurgeryClinical Vision ScienceOcular DiseaseOphthalmic SurgeryCataract ManagementCataract SurgeryLens DiseaseLens SurgeryMacular DegenerationOcular ImplantsAcuity In CataractAge-related Macular Degeneration ImplantsLens DesignPostoperative Visual AcuityPostoperative Visual Outcome AssessmentVisual Acuity ImprovementVisual Impairment Due To Cataract
The teledioptric lens implant, with a high-minus central zone, was evaluated on an optical bench and implanted in the eyes of 50 patients with both cataract and age-related macular degeneration. In vitro testing demonstrated slightly reduced contrast and modulation-transfer-function as compared with a standard intraocular lens. The lens implant functions ideally with a 1.9-millimeter diameter on the high-minus portion. Lens implantation was proven safe in the clinical study. When used with spectacles as a part of a teledioptric system, the resulting visual field was 2.6 times larger than that achieved with an external telescope. Postoperative visual acuity initially improved in more than 65% of patients. After 1 year, due to disease progression, only 20% had improved acuity.
4
Harold A. Kahn, Howard M. Leibowitz, James P. Ganley et al. · American Journal of Epidemiology · 1977 · 399 citations
Cataracts and Macular Degeneration in Older Americans
Ronald Klein · Archives of Ophthalmology · 1982 · 274 citations
Visual impairment of age‐related macular degeneration
Troels Vinding · Acta Ophthalmologica · 1990 · 45 citations