Temperature-guided high and very high-power short duration ablation for atrial fibrillation treatment: the peQasus multicentre study
13
Citations
30
References
2024
Year
In this large multicentre study, temperature-controlled HPSD and vHPSD ablation via a novel ablation catheter provides safe and effective PVI with a relatively short procedure duration. Despite a shorter procedure time, no differences in terms of safety and freedom from arrhythmia recurrence were found irrespective of utilizing vHPSD or the hybrid approach.