Asian Spine Journal · 2024 · 11 citations · 12 references
MedicineOrthopedic SurgerySurgeryMedical RoboticsOrthopedicsSpine SurgeryFluoroscopy-guided ProceduresIntraoperative NavigationMinimally Invasive Spine SurgeryRobot-assisted SurgerySurgical RoboticsAnchorsIntraoperative Image GuidanceNavigation-guided Spine SurgeryNavigation-guided SurgeryPlacementRobotic-assisted Spinal SurgeryRobotic-assisted Spine SurgeryTomography
STUDY DESIGN: Prospective single-center study. PURPOSE: To compare the accuracy of pedicle screws placed by freehand and under fluoroscopy and robotic assistance with intraoperative image acquisition. OVERVIEW OF LITERATURE: Pedicle screws are the most commonly used spinal anchors owing to their ability to stabilize all three spinal columns. Various techniques such as freehand, fluoroscopy-assisted, and navigation-assisted pedicle screw placements have been used with varying degrees of accuracy. Most studies on robotic-assisted pedicle screw placement have utilized preoperatively acquired computed tomography scans. To our knowledge, this is the only study in the literature that compared freehand with fluoroscopy-guided and robotic-assisted pedicle screw insertion with freehand and fluoroscopy. METHODS: In this prospective study, a total of 1,120 pedicle screws were placed in the freehand group (n=175), 1,250 in the fluoroscopyassisted group (n=172), and 1,225 in the robotic-assisted group (n=180). Surgical parameters and screw accuracy were analyzed between the three groups. The preoperative plan overlapped with the postoperative O-arm scan to determine if the screws were executed as planned. RESULTS: The frequency of clinically acceptable screw placement (Gertzbein-Robbins grades A and B) in the freehand, fluoroscopy-assisted, and robotic-assisted groups were 97.7%, 98.6%, and 99.34%, respectively. With robotic assistance, an experience-neutralizing effect implied that surgeons with varying levels of experience achieved comparable pedicle screw accuracy, blood loss, O-arm time, robot time, and time per screw. No significant difference in these parameters was found between surgeries commencing before and after 2 PM. No significant differences were noted between the planned and executed screw trajectories in the robotic-assisted group irrespective of surgical experience. CONCLUSIONS: The third-generation robotic-assisted pedicle screw placement system used in conjunction with intraoperative threedimensional O-arm imaging consistently demonstrates safe and accurate screw placement with an experience-neutralizing effect.
12
Accuracy of Pedicular Screw Placement In Vivo
Stanley D. Gertzbein, STEPHEN E. ROBBINS · Spine · 1990 · 1.4K citations
Free Hand Pedicle Screw Placement in the Thoracic Spine: Is it Safe?
Yongjung J. Kim, Lawrence G. Lenke, Keith H. Bridwell et al. · Spine · 2004 · 706 citations
Bawarjan Schatlo, Granit Molliqaj, Victor Cuvinciuc et al. · Journal of Neurosurgery Spine · 2014 · 230 citations