IEEE Sensors Journal · 2024 · 11 citations · 32 references
EngineeringBioengineeringBiomedical EngineeringBiosensorsPhotonicsSensorsMicrofluidic SystemsMicrofluidicsOptical BiosensorsOptical FibersOptical SensingOptical SensorsFabricationGlucose BiosensingGlucose ConcentrationGlucose MonitoringGlucose SensingGlucose SensorsOptical Fiber SensorsPolymethyl Methacrylate
This study introduces a successfully manufactured U-shaped optical fiber sensor integrated with a microfluidics system to detect varying glucose concentrations. The outlines the fabrication process of the microfluidics system, accomplished using a CO2 laser on polymethyl methacrylate (PMMA) material. Through three cycles of experiments involving different glucose concentrations, the results indicate a consistent trend in both resonance wavelength and transmission loss, showing an increase as glucose concentration rises from 5% to 50%. The system demonstrates a high wavelength sensitivity coefficient of 0.114 nm/% with R <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sup> = 0.95 and a notable transmission loss sensitivity of 0.09 dB/% and R <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sup> = 0.94. This advanced system proves to be effective in achieving rapid, sensitive, and cost-efficient glucose monitoring. Consequently, the microfluidics system is deemed suitable for its intended purpose.
32
Hesham Azizgolshani, Jonathan Coppeta, Else M. Vedula et al. · Lab on a Chip · 2021 · 219 citations · Full text
Recent progress of microfluidic reactors for biomedical applications
Huanhuan Shi, Kaixuan Nie, Bo Dong et al. · Chemical Engineering Journal · 2018 · 202 citations