Design, manufacturing and testing of a green non-isocyanate polyurethane prosthetic heart valve

Sofia Ferreira Melo, Alicia Nondonfaz, Abdelhafid Aqil, Anna Pierrard, Alexia Hulin, Céline Delierneux, Bartosz Ditkowski, Maxime Gustin, Maxime Legrand, Bibian M. E. Tullemans,

Biomaterials Science · 2024 · 20 citations · 53 references

DOIFull text

Open access

Concepts

Abstract

, and used to produce aortic valves. PHU elastomers reinforced with a polyester mesh show mechanical properties similar to native valve leaflets. These NIPUs do not cause hemolysis. Interestingly, both platelet adhesion and contact activation-induced coagulation are strongly reduced on NIPU surfaces, indicating low thrombogenicity. Fibroblasts and endothelial cells maintain normal growth and shape after indirect contact with NIPUs. Fluid-structure interaction (FSI) allows modeling of the ideal valve design, with minimal shear stress on the leaflets. Injection-molded valves are tested in a pulse duplicator and show ISO-compliant hydrodynamic performance, comparable to clinically-used bioprostheses. Poly(tetrahydrofuran) (PTHF)-NIPU patches do not show any evidence of calcification over a period of 8 weeks. NIPUs are promising sustainable biomaterials for the manufacturing of improved prosthetic valves with low thrombogenicity.