Concepedia

Publication | Open Access

Three-dimensional intracardiac echocardiography for left atrial appendage sizing and percutaneous occlusion guidance

33

Citations

18

References

2023

Year

Abstract

Intracardiac echocardiography-guided LAAO showed a very high success, with no major adverse events. A very high level of agreement for LAA sizing was found between pre-procedural TEE and periprocedural 3D-ICE. 3D-ICE performed significantly better than 2D-ICE for FLX size selection and may provide better guidance during device deployment.

References

YearCitations